PURPOSE
To lay down a procedure for the Calibration of the FTIR Spectrophotometer, Model: IRSpirit-X, Make: Shimadzu
SCOPE
This SOP applies to the Calibration of the FTIR spectrophotometer, with the model IRSpirit-X and the make Shimadzu.
RESPONSIBILITY
- QC Analyst: Calibration of an instrument as per SOP.
- Senior Analyst: Reviewing of recorded data and implementation of the SOP.
- Manager QAD/QCD: Approval and responsible for implementing the SOP.
Handling of polystyrene film
- Polystyrene film thickness should be approximately 35µm.
- Ensure that the polystyrene film is within its validity.
- After completion of calibration store the polystyrene film in a self-sealed polyethylene bag/ box and keep it in a desiccator.
- Label the polystyrene film with the following information from its certificate of analysis.
|
Polystyrene
film |
|
|
Lot No. |
|
|
Calibrated on |
|
|
Valid till |
|
|
Sign & Date |
|
Calibration Procedure
FTIR is calibrated by using polystyrene film for the following parameters,
- Verification of wave number accuracy.
- Control of Resolution performance of the instrument.
- Click on the ‘Macro’ tab in the LabSolutions IR launcher, following window appears.
- Select ‘Measurement’, and the following screen appears.
- Click on ‘OK’, and the following screen appears.
- Put the polystyrene film in the sample compartment and click on ‘OK’.
- When the measurement is finished the calculation for inspection is done automatically. The validation program inspects 5 tests. When all tests pass, the validation is passed. The judgment dialog box shows ‘Pass’ and a validation report is printed out.
- Then close the program by clicking on ‘close’ / ‘Program end’.
- Repeat the same procedure for ‘EP Validation’.
- The spectrum shall give the wave number accuracy at the wavelength mentioned below. Record the observation in the Calibration report as per ‘CALIBRATION REPORT OF FTIR SPECTROPHOTOMETER’ Annexure 01.
- Similarly, check the performance of ATR (Attenuated Total Reflection) mode by selecting ‘ATR’ in the EP Validation parameter setting.
- Acceptance criteria: The wave number accuracy obtained with the polystyrene film shall be within the range specified in the Calibration compilation report as per ‘Calibration Compilation Report for FTIR’.
- Verification of Wave Number for Transmission mode as per the following table.
|
Wave
Number |
Tolerance
Limit |
Acceptance
Criteria |
|
3060.0 |
± 1.0 cm-1 |
3059.0 – 3061.0 |
|
2849.5 |
2848.5 – 2850.5 |
|
|
1942.9 |
1941.9 – 1943.9 |
|
|
1601.2 |
1600.2 – 1602.2 |
|
|
1583.0 |
1582.0 -1584.0 |
|
|
1154.5 |
1153.5 – 1155.5 |
|
|
1028.3 |
1027.3 – 1029.3 |
|
|
906.6 |
905.6 – 907.6 |
- Verification of Wave Number for ATR mode as per the following table.
|
Wave
Number |
Tolerance
Limit |
Acceptance
Criteria |
|
3059.7 |
± 1.0 cm-1 |
3058.7 – 3060.7 |
|
1601.0 |
1600.0 – 1602.0 |
|
|
1027.7 |
1026.7 – 1028.7 |
|
|
906.1 |
905.1 – 907.1 |
- Control of Resolution Parameter by Absorbance as per the following table.
|
Absorbance
difference between (cm-1) |
Tolerance
Limit |
|
2870.0 – 2849.5 |
NLT 0.33 ABS |
|
1589.0 – 1583.0 |
NLT 0.08 ABS |
- Frequency of Calibration: Once in 6 months or after any major maintenance job.
- Calibrate the instrument on or before the calibration due date.
- After completion of the calibration activity, affix the calibration status label on the instrument.
- If any calibration result is found to not meet the acceptance criteria specified in SOP, this shall be considered out of calibration.
- If calibration complies, update an instrument's current calibration status label with all relevant details indicating the calibration status as ‘PASSED’.
ANNEXURES
Calibration Report for FTIR Spectrophotometer
REVISION HISTORY
Nil









0 Comments