OBJECTIVE
To provide standard instructions for the operation of the Malvern Zetasizer (Nano-ZS-ZEN3600)
SCOPE
The operation of the particle size analyzer is to measure particle size, molecular weight, and zeta potential.
RESPONSIBILITY
- QC analysts and related personnel are responsible for the implication of this procedure.
- QA head is responsible for regular and required amendments in the Standard Operating Procedure.
PRECAUTIONS
- Make sure to switch off the instrument when not in use.
- Cells should be cleaned prior to use. All sizes and Zeta potential cells should be rinsed/ cleaned with filtered dispersant before use.
- The cell should be filled slowly to avoid air bubbles. Ultrasonication can be used to remove air bubbles, but only if the sample is suitable for use with ultrasonics.
- When using a syringe for the dispersant, never use the first few drops from the syringe to avoid contamination of the dispersant by residual dust particles.
- Hold the cuvette from the top to prevent tampering with the measurements, and wipe the sides with clean tissue.
PROCEDURE
- Turn ON the instrument and wait for 1 hour for the laser to stabilize.
- Start the Zetasizer software.
- Choose the appropriate one for the measurement. Organic solutions can only be measured in a glass cuvette.
- Fill the cuvette to a depth of approximately 1 cm.
- Make a new measurement file, then select Measure → Manual.
- Click on the measurement type to choose the experiment type.
- Click on the sample and choose the name, and write any notes as needed.
- Click on Sample Material and choose your sample material from the list.
- You may add a new material if the refractive index is known.
- Choose a dispersant or add a new one as needed.
- Temperature: at least 120s equilibrium time should be allowed since the room is very cold.
- Cell type: choose a category and model.
- Measurement: duration and repetitions may be adjusted.
- Press OK to accept the measurement settings.
- Load the sample and press Start.
- To view the data, select a measurement and look through the tabs (e.g., intensity or volume PSD).
- On a particular tab, select Edit → Copy Size Values/Graph.
- Then the data can be transferred to an Excel spreadsheet.
- Turn off the instrument.
CLEANING AND MAINTENANCE
- Switch OFF the instrument.
- Switch OFF the mains for the power supply.
- Clean the instrument with a soft de-dusting cloth.
- Care must be taken that samples are not split on the instrument or in the cuvette chamber.