OBJECTIVE
To lay down a Systematic procedure for the operation and calibration of the Dissolution Test Apparatus (Model: VDA-8D).
SCOPE
This SOP is a guideline for the operation and calibration of the Dissolution Test Apparatus (Model: VDA-8D).
RESPONSIBILITY
Q.C. Officer: To follow the procedure for operation of Dissolution Test Apparatus as per the SOP, as well as to carry out calibration of Dissolution Test Apparatus.
Q.C. In-charge: To check that the operation and calibration procedure ofthe Dissolution Test Apparatus is being followed.
Head of the Department
PRE-CHECK
Ensure that the apparatus is clean and the plug is connected to the Electricity Supply.
OPERATING PROCEDURE
- By removing one of the vessels, fill the acrylic tank with distilled water up to the level mark. Then fit the dissolution vessel back in its place.
- Choose either six or eight stirring shafts with the blades or baskets as per the requirements of the test.
- Insert these shafts from the top plate of the instrument, which is having through hole and is marked as 2, 3, 4, 5 to the front row and 1, 8, 7, 6 to the second row.
- The stirrer shafts are marked as 1-2-3-4-5-6-7-8. Insert stirrer shaft no. 1 into spindle no. 1 of stirrer platform as so on.
- Fill the dissolution vessels with the dissolution media up to the mark, 900ml, or 1000ml or as per the requirement.
- This indicates that the temperature of the bath has reached 38℃. Still the test cannot be started as the temperature of the vessel yet to attain 37℃.
- The Stirrer Platform will automatically go down to its fixed position.
- Select ‘F1’ from the keyboard or select ‘Menu Start’ key from the instrument. Then select ‘TEST’.
- Select Operator Name, enter password, then select assembly type i.e. paddle type or basket type.
- Select the product whose test is to be carried out and press the ‘ENTER’ key twice.
- After the vessel reaches the set temperature, press the ‘ENTER’ key to start the machine.
- The machine will stop after it has travelled the set time.
- Use the ‘Platform Up’ and ‘Platform Down’ key from the keyboard for up and down movement of the Stirrer Platform.
- If necessary, Use ‘STR ON’ and ‘STR OFF’ key for the Stirrer on and off.
- For setting the parameters for a product
- Press ‘F2’ key from the keyboard.
- Enter ‘1, 2, 3, 4, 5, 6’ as the password.
- Select the empty product. Enter all the required parameters (i.e. Product Name, Temperature, RPM, Alarm Time and Test Time) and press ‘ENTER’.
- Go back to home page by pressing ‘Esc’ Key.
- Carry out the test from the above point onward.
- Press ‘Start’ key. The display show,
- Select Synchronous and press ‘ENTER’.
- The display shows Sample Vol. and Test No. Press any key to continue.
- Enter the required sample volume.
- Enter the required test no. (Test no. is the no. of Alarm set). Press ‘ENTER’ two times.
- The display will show Select position to Start. Select Home position and then press enter. The test starts.
CALIBRATION
Temperature inside the vessel
- After the temperature has been attained to the defined limit, record the temperature of the medium at different time intervals (0, 10, 20 min.) with a calibrated thermometer.
- Compare the temperature of the calibrated thermometer with the set temperature.
- Limit (USP Specification): 37±0.5ºC.
Rotation of Stirring Shaft
- Set the rotation of the stirring shaft to 50, 75, 100 RPM as per the procedure defined.
- Use a Tachometer or manually count the number of rotations against time (per minute).
- Compare the observed rotation with the set rotation of the apparatus.
- Limit (USP Specification): 4% of set RPM i.e. 48 RPM – 52 RPM (if set RPM is 50).
Distance between the inside bottom of the vessel and the basket/paddle.
- Use a calibrated measuring scale to measure the distance between the inside bottom of the vessel and the basket/paddle.
- Place the measuring scale to the adjacent side of the basket/paddle and measure the distance between the basket/paddle and the bottom of the vessel.
- Limit (USP Specification): 25±2 mm.
- Record all the data.
- If it does not meet the limit, inform to the head of the department for the necessary corrective action.
Calibration Frequency: Quarterly
ABBREVIATIONS
SOP: Standard Operating Procedure
USP: United States Pharmacopeia
RPM: Revolution Per Minute
REFERENCES
User Manual, USP
ANNEXURE
Dissolution Test Apparatus Log Sheet
Dissolution Test Apparatus Calibration Report
REVISION HISTORY
Nil