OBJECTIVE
To calibrate the equipment for reliable and accurate results.
SCOPE
This procedure applies to the Polarimeter, installed in Pharmainfo Quality Control Lab.
RESPONSIBILITY
Doing: Technical Assistant
Checking: Executive /Manager
Head of the Department
PROCEDURE
Frequency: Once in a three month
- Operate the instrument as per SOP.
- Prepare 10%, 20%, 30%, 40% and 50% w/v aqueous solution of AR-grade sucrose (previously dried) in purified water.
- Note the temperature of the solution and determine the optical rotation three times by changing the solution in the tube and record the mean.
- Calculator corrected Optical Rotation at 20℃ (a 20) using the following formula.
a20 = Optical rotation to be corrected at 20oC
at = Optical rotation to be corrected at toC
t = Temperature measured by sensor
k = Temperature correction factor (-0.00045)
- Record the value in Annexure I.