SOP for Calibration of Brookfield Digital Polarimeter

OBJECTIVE
To calibrate the equipment for reliable and accurate results.

SCOPE
This procedure applies to the Polarimeter, installed in Pharmainfo Quality Control Lab.

RESPONSIBILITY
Doing: Technical Assistant
Checking: Executive /Manager
Head of the Department

PROCEDURE
Frequency: Once in a three month
  • Operate the instrument as per SOP.
  • Prepare 10%, 20%, 30%, 40% and 50% w/v aqueous solution of AR-grade sucrose (previously dried) in purified water.
  • Note the temperature of the solution and determine the optical rotation three times by changing the solution in the tube and record the mean.


  • Calculator corrected Optical Rotation at 20℃ (a 20) using the following formula.

Where
a20 = Optical rotation to be corrected at 20oC
at = Optical rotation to be corrected at toC
t = Temperature measured by sensor
k = Temperature correction factor (-0.00045)

  • Record the value in Annexure I.
ABBREVIAITON
% = Percentage
w/v= weight by volume


ANNEXURE

REVISION HISTORY
Nil